Retro-Fit Manual Puritan-BennettSeries A, Series 73 and Litharged Outlet www.amico.com
10 Amico Pipeline1. Insert the latch valve assembly in the assembled ller frame and face plate.2. Align the assembled parts with the mounting b
www.amico.com 11PARTS BREAKDOWNA B* 1 O-CPBF-ZZ-L-GAS LATCH VALVE ASSEMBLY* 2 H-MPF-0612 #6-32 X 3/4"3 O-TRIM-M325C Chrome Trim Plate 3.
APE PB RETROFIT MANUAL DEC 2011www.amico.comAmico Pipeline | www.amico.com 85 Fulton Way, Richmond HillOntario, L4B 2N4, Canada71 East Industry Court
2 Amico PipelineTABLE OF CONTENTS Introduction 3Components Required 4Preparing Outlets for Installation 5Removing Litharged Primary Valve and C
www.amico.com 3INTRODUCTIONThe Retro-Fit process allows you to update and convert your old Puritan-Bennett outlets to new Amico outlets. This
4 Amico PipelineRETROFIT COMPONENTS REQUIREDO-CSS-ZZ-L-G-W SS = denes Series: ZZ = denes Latch Valve Assembly Style: L = denes Language: G =
www.amico.com 5PREPARING OUTLETS FOR RETROFIT INSTALLATIONWARNING: Oxygen, while not ammable, supports rapid combustion of ammable substan
6 Amico PipelineThe valve connector is held in place by the existing oxygen compatible compound that must be heated in order to loosen parts for r
www.amico.com 7Place the new O-ring (supplied with the kit) on the end of the connector. Carefully hand thread the connector into the rough-i
8 Amico Pipeline1. Be certain the rough-in and outlet box are free of dirt and debris.2. Make certain the new O-ring seal is in place in the gro
www.amico.com 9The indexing plate provides gas specic pin indexing and color coded labeling for upgraded outlets, to prevent cross assembly